SCALE OPERATION
CALIBRATION
Bar Code Calibration Procedure without Remote Display*
13. Calibration is now complete. The scanner will automatically restart, and beep one time, in 5 seconds.
14. In order to use the scale in the normal operating mode, the unit must be powered down and the scale program switch
returned to the run position.
Figure 61.
See power source caution statement on page 8 of this manual.
15. The unit’s calibration must now be verified as required by state and/or local Weights and Measures regulations
(starting on page 46).
* These procedures are for MS2320 StratosH Scanner/Avery Scale models only.